The Precision:
Iterm Range of sample:SD
HB:4.5g/dL-10.0g/dLSD≤0.8g/dL
10.1g/dL-25.6g/dLCV≤8%
Specifications:
Test Time≤ 15 seconds
Measurement Range:
4.5g/dL-25.6 g/dL
45g/L-256 g/L
The Accuracy:
Iterm Range of sample:Bias
HB:4.5g/dL-10.0g/dL≤±1g/dL
10.1g/dL-25.6g/dL≤±10%
Sample:Whole bl-oo-d(capillary bl-oo-d and venous whole bl-oo-d)
Volume:10 μL
Package Contents:
50x Strips,50xLancets (The Test Strips only can be used on the same model of test meter BHM-101/102 Hemoglobin Meter)